National Patient Safety Alert

 

National Patient Safety Alert: Medical beds, trolleys, bed rails, bed grab handles and lateral turning devices: risk of death from entrapment or falls

Earlier this week, the MHRA sent an alert following a number of received reports of deaths and serious injuries from entrapment or falls relating to medical beds, bed rails, trolleys, bariatric beds, lateral turning devices and bed grab handles. 

This alert is for action by: All those responsible for the use, purchase, prescription and maintenance of medical beds, trolleys, bed rails, bed grab handles and lateral turning devices including all Acute and Community healthcare organisations, care homes, equipment providers, Occupational Therapists and early intervention teams. 

According to investigations, deaths were found to involve factors including inadequate risk assessment, maintenance issues and children and adults of small stature using beds which are designed for use by adults with typical body dimensions. 

Other risk factors (such as inappropriate use or incompatibility) are included in the MHRA’s updated guidance on the management and safe use of bed rails and should be considered as part of an appropriate risk assessment.  

Actions required 

When: Begin as soon as possible and complete by 1 March 2024 

  1. Update your organisation’s policies and procedures on procurement, provision, prescribing, servicing and maintenance of these devices in line with the MHRA’s updated guidance on the management and safe use of bed rails. 

  1. Develop a plan for all applicable staff to have training relevant to their role within the next 12 months with regular updates. All training should be recorded. 

  1. Review the medical device management system (inventory/database) for your organisation or third-party provider for devices within your organisation, including those which have been provided to a community setting (for example, the patient’s own home). Keep this system up to date. 

  1. Implement maintenance and servicing schedules for the devices in the inventory/database, in line with the manufacturer’s instructions for use and/or service manual. Prioritise devices which have not had regular maintenance and servicing. If this is outsourced, compliance with the schedule should be monitored. 

  1. Review patients who are children or adults with atypical anatomy as a priority. Ensure the equipment they have been provided with is compliant with BS EN 50637:2017 unless there is a reason for using a non-compliant bed. Record this on the risk assessment and put in place measures to reduce entrapment risks as far as possible. 

  1. Review all patients who are currently provided with bed rails or bed grab handles to ensure there is a documented up-to-date risk assessment. Complete risk assessments for patients where this has not already been done and for each patient who is provided with bed rails or bed grab handles. 

  1. Implement systems to update risk assessments where the equipment or the patient’s clinical condition has changed (for example, reduction/improvement in weight or mobility), and also at regular intervals. 

You can read the alert in full on the Government Website